Specifications
Measurement Range
0.01 –1100 NTU/FTU
Accuracy
±2% of reading or 0.01 NTU (0-500 NTU)
±3% of reading (500-1100 NTU)
Resolution
0.01 NTU on low readings
Regulatory Compliance Compliant to ISO 7027: Water Quality – Determination of
Turbidity (Nephelometric Method)
Response Time 6 - 16 seconds
Display 7 Segment Liquid Crystal Display
Operating Temperature Range
0°C – 50°C (32°F – 122°F)
Sample Temperature Range 0°C – 50°C (32°F – 122°F)
Relative Humidity 0-90% Non Condensing
Power Supply 4 – AAA Batteries (life is approximately 1000 readings using
standard alkaline batteries)
Certification CE
Insulation Rating Pollution Degree 2
Approximate Dimensions Instrument: 7 cm x 16.5 cm x 4.8 cm (2 ¾” x 6 ½” x 1 7/8”)
Instrument in Minilab Case: 25 cm x 21.6 cm x 6.3 cm
(10” x 8 1/2” x 2 1/2”)
Enclosure Rating:
NEMA 4X
Designed to meet the specifications of IP67
Shipping Weight: 1.22 kg (2.70 lbs)
Shipping Dimensions: 28 cm x 30.5 cm x 7.6 cm (11” x 12” x 3”)
Warranty 1 Year from date of shipment
Ref. 22666 (12/04)